071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
9-15
Technical Specifications
Electromagnetic Compatibility Statement
Table 9-3
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump
should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The pump uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The pump is suitable for use in all establishments, other than
domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Table 9-4
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV contact (1)
Floors should be wood, concrete, or ceramic tile. If the
floors are covered with synthetic material, the relative
humidity should be at least 30%.
± 8 kV air
± 15 kV air (1)
Pump may stop infusing and alarm at or above ± 15kV
air.
Electrical fast
transient burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1kV for input/output
lines
± 2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Input/output lines are not used in the patient area.
Surge
IEC 61000-4-5
± 1kV differential
mode
± 2 kV common mode
± 1kV differential
mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...