COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
9-13
Technical Specifications
Accuracy Tests
3. Bolus Volume Released After Downstream Occlusions
Are Corrected
4. Maximum Volume Under Single Fault Condition
1157 microliters ±5% delivered in 2 seconds.
Rate
Occlusion Alarm
Pressure Setting
Typical
Bolus
Volume
Maximum
Bolus
Volume*
1 mL/hr
Minimum 103 mmHg (2 psig)
Moderate 258 mmHg (5 psig)
Maximum 465 mmHg (9 psig)
0.1 mLs
0.1 mLs
0.2 mLs
0.1 mLs
0.2 mLs
0.3 mLs
25 mL/hr
Minimum 207 mmHg (4 psig)
Moderate 414 mmHg (8 psig)
Maximum 620 mmHg (12 psig)
0.1 mLs
0.2 mLs
0.3 mLs
0.2 mLs
0.2 mLs
0.3 mLs
* At least 90% of the observed values (95% confidence) will lie below the limits
shown for the indicated settings.