COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
8-13
Troubleshooting
Troubleshooting Alarms
Table 8-1
Troubleshooting Alarm Messages
Alarm Message
Cause
Recommended Action
X - Close Regulating
Clamp
PATIENT---->>>>
One of the following occurred
while an administration set was
loaded:
• The
Open
key was pressed.
• The Manual Tube Release was
opened.
•
Close the regulating clamp
on the
administration set
and remove the keyed slide
clamp from the slot.
• Reload the set, if desired.
Damaged Battery! Service
Now
(no pump module alert message)
This message occurs as an alarm
if the pump detects one or more
of the following conditions
during power up:
• The battery's history data was
cleared inadvertently because of
an external event.
• The battery voltage falls below
10.4 volts.
• The battery has been charged
and discharged more than 200
times.
When a
Damaged Battery! Service Now
alarm occurs during power on, the pump
cannot be used. See “About the Damaged
Battery Alarm,” 8-8 for more information.
X - Downstream Occlusion
DWN OCCL/label
A closed clamp, stopcock,
clogged filter or other occlusion
is preventing fluid flow between
the pump and patient.
• Correct the problem causing the occlusion.
• Restart the infusion:
• for single channel pumps: press the
Primary
or
Secondary
soft key to
access the appropriate programming
screen and then press the
START
key.
• for triple channel pumps: press the
appropriate
Channel Select
key to
access the appropriate programming
screen and then press the
START
key.
Note:
When the pump is configured with the
Auto Restart feature on, the pump channel can
automatically restart if the occlusion is removed
within one minute after detection. If any pump key
is pressed during a
Downstream Occlusion
alarm, Auto Restart will be disabled.
X - Incomplete Primary
Program
X - Incomplete Secondary
Program
STOPPED
The
START
key was pressed
prior to completing
programming.
Enter the missing parameter value(s) and press
the
Confirm Primary
or
Confirm Secondary
soft key followed by the
START
key.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...