6
Options Specific to PERSONALITY Feature Sets
6-4
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Options Specific to PERSONALITY Feature Sets
The options listed in Table 6-3 can be set independently for each
PERSONALITY feature set.
Table 6-3
Options Specific to a PERSONALITY Feature Set
Option
Available Settings
Factory
Settings
Volume-Time Infusion
Mode
Available (Enable)
Unavailable (Disable)
Enabled
Delay Start
Available (Enable)
Unavailable (Disable)
Disabled
Secondary Infusion Mode
Available (Enable)
Unavailable (Disable)
Enabled
Dose Infusion Modes
Available (Enable)
Unavailable (Disable)
Enabled
Individual dose formulas
mg/hr
mg/min
mcg/hr
mcg/min
units/hr
mEq/hr
mg/kg/hr
mg/kg/min
mcg/kg/hr
mcg/kg/min
units/kg/hr
mEq/kg/hr
(When Dose Infusion Modes is enabled, individual modes can be
enabled or disabled.)
Enabled
Enabled
Enabled
Enabled
Enabled
Disabled
Enabled
Enabled
Enabled
Enabled
Enabled
Disabled
Pump Channel Message
Display
• Rate (mL/hr), used if no label is selected
• Time Remaining (Before KVO)
• Label (the Label Library feature and individual labels must be
enabled)
• Volume Infused (mL)
Rate
Prime Feature
Available (Enable)
Unavailable (Disable)
(Enables priming for administration sets difficult to
gravity-prime.)
Enabled
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...