COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-3
Quick Reference Guide
Power On and PERSONALITY Feature Set Selection
Power On and PERSONALITY Feature Set Selection
Note:
The COLLEAGUE and COLLEAGUE 3 pumps have not been
evaluated for use in care areas other than those listed on
1.
Press the
ON/OFF CHARGE
key to turn the pump ON. The Main
Display prompts to perform a speaker test, which helps ensure
alarms and alerts are audible and the volume level is appropriate for
the care area.
2.
Press and hold the
Speaker Test
soft key until
Yes
and
No
soft keys
are displayed (Figure 11-1). The pump produces sound for as long as
the
Speaker Test
soft key is pressed.
3.
Do one of the following:
If the tone is heard, press
Yes
. The pump completes self-test and
displays the Power On screen.
If the tone is not heard, even after adjusting the volume control,
press
No
, then press
No
again when prompted to confirm. Do not
use the pump.
If the continuous tone is not heard during the speaker test, alarms and
alerts may not be audible during operation. Do not use the pump.
Send the pump to service.
4.
To clear previously programmed information and Volume History,
press the
New Patient
soft key.
5.
To select a PERSONALITY feature set, press the
Change
Personality
soft key. A list of available PERSONALITY feature
sets is displayed.
Figure 11-1 Speaker Test
Screen
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...