COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
4-63
Operating Instructions
Adding or Changing a Label Line During an Infusion
Adding or Changing a Label Line During an Infusion
If the Label Library feature is enabled, use the procedure below to select
an informational label during an infusion.
1.
From the Main Display:
2.
Press the
Change Mode
soft key. The Programming Modes menu is
3.
Highlight
Label Line
(under
Functions
) using the
×Ø
keys, then
press the
Select
soft key.
A list of labels and their abbreviations is displayed as shown in
Figure 4-29. If the list consists of more than one screen, use the
Page
Up
and
Page Down
soft keys to view the next screen of labels.
4.
Highlight the label to be selected using the
×Ø
keys, then press the
Select
soft key. When the
Select
soft key is pressed, the
Programming screen is displayed, showing the selected label.
5.
Press the
START
key to resume the infusion.
Viewing COLLEAGUE GUARDIAN Limits During an Infusion
Use this procedure to view the COLLEAGUE GUARDIAN
programming limits for the infusion currently shown on the display.
Figure 4-91 Programming Modes
Menu
For single channel pumps:
For triple channel pumps:
press the
STOP
key, then press
the
Primary
soft key.
press the
STOP
key on the
desired pump channel, then
press the
Channel Select
key.
Note:
COLLEAGUE
GUARDIAN
Labels
Labels configured using the COLLEAGUE GUARDIAN
feature do not appear in the label list.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...