Programming a COLLEAGUE GUARDIAN Dose Mode Infusion
11-10
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Press
Accept Rate
(
×
key) to accept the out-of limits flow rate
and continue with the infusion as programmed.
7.
Press the
START
key to begin the infusion.
COLLEAGUE GUARDIAN infusions are indicated by the mortar
and pestle icon
next to the label on the Main Display screen.
Programming a COLLEAGUE GUARDIAN Dose Mode Infusion
1.
For single channel pumps: press the
Primary
soft key.
For triple channel pumps: press the desired
Channel Select
key.
2.
Press the
Colleague Guardian
soft key. The labels for which
COLLEAGUE GUARDIAN limits have been defined are displayed
in a pop-up window (Figure 11-8).
3.
Use the
×Ø
keys to highlight the desired label, then press the
Select
soft key.
The programming mode changes to the mode configured for the
selected label, and the
Drug Amount, Diluent Volume
, and
Concentration
fields are filled with the defined values.
For non-weight-based modes, the
Volume To Be Infused
field
is filled with the standard Diluent Volume (Figure 11-9).
Figure 11-8 Colleague Guardian
Label List
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...