COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-11
Quick Reference Guide
Programming a COLLEAGUE GUARDIAN Dose Mode Infusion
For weight-based modes, the
Weight
11-10). If a default dose has been configured, it appears in the
Dose
field.
4.
For weight-based modes, enter patient weight using the numeric
keypad. The pump calculates the values for the remaining fields.
Depending on how the pump has been configured at the facility, the
kg
or
lbs
field may not be available for data entry. Fields not
available for data entry appear as shaded.
For small patients, weight can be entered in grams (or ounces) if
appropriate. To change weight units, highlight the
Weight
field,
press the
Units
soft key to display the weight units list, highlight the
desired weight unit, then press the
Select
soft key.
5.
(Optional) To view the limits for the label, press the
View Limits
soft
key. The Limits Display pop-up displays the preconfigured range
limits for rate, dose, or concentration, whichever is appropriate for
that label (Figure 11-11). Press the
Done
soft key to close the
pop-up.
Figure 11-9 Non-Weight-Based
Programming Screen
Figure 11-10 Weight-Based
Programming Screen
Figure 11-11 View Limits Pop-up
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...