Programming a Delay Start Infusion
4-58
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Powering On with Delay Start Infusions
If the pump was powered off while in Delay Start Infusion mode, and
the pump is powered on within the 5-hour memory retention period, the
pending Delay Start infusion is cancelled and the Start At time is
cleared. Delay Start is still the active infusion mode for the channel, but
the infusion has not been completely programmed.
The Delay Start mode will be cancelled by using the
Change
Personality
pop-up and pressing the
Select
soft key or by pressing the
New Patient
soft key.
Note:
Delay Start and
Standby
A completely programmed Delay Start infusion may not
be placed into Standby. If the pump channel is placed
into Standby with a completely programmed Delay
Start infusion, the Start At time is cleared.
Note:
Manual Tube
Release
If a Manual Tube Release alarm occurs on a pump
channel with a completely programmed Delay Start
infusion, the Delay Start infusion is cancelled and the
Start At time is cleared.