071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
7-5
Maintenance and Storage
Maintenance
Optimizing Battery Service Life
Ensuring the following procedures are a part of routine device use can
optimize battery service life:
Connect pumps to AC power at all times except in the event of
AC power loss or short-term portable operation.
Store pumps plugged into AC power to maintain the battery
charge whenever possible.
If the batteries have been discharged, do not store the pump
without first fully charging the batteries. The batteries may be
permanently damaged if the pump is stored unplugged with
discharged batteries for more than two weeks.
To help prolong battery life, never store the pump unplugged. See
“Storage,” 7-6 for more information.
Notify Central Supply or other appropriate departments as soon
as a pump is removed from patient use so that it can be cleaned
and the batteries can be recharged.
Recharge a pump that is in the
DEPLETED Battery - Plug In
Now
alert condition or the
NO Battery - Plug In Now
alarm
condition for a minimum of 12 uninterrupted hours or until the
battery operating time displayed under the battery charge icon in
the upper left part of the screen is 3h 15m (for triple channel
pumps) or 4h 00m (for single channel pumps). A complete charge
may take longer than 12 hours.
Charging the Batteries
immediately. Do not use the pump on battery power until the batteries
have been fully recharged. Charge the batteries for at least
12 uninterrupted hours or until the battery operating time displayed
under the battery charge icon in the upper left part of the screen is
3h 15m (for triple channel pumps) or 4h 00m (for single channel
pumps). A complete charge may take longer than 12 hours.
The batteries are charging whenever the pump is plugged into a
100-120 VAC, 50/60 Hz or 220-240 VAC 50/60 Hz outlet, regardless of
whether the pump is on or off. Whenever the batteries are charging, the
Plug icon is lit. Store the pump plugged in to maintain batteries at full
charge. See “Storage,” 7-6 and the
COLLEAGUE Pump Global Service
Manual
for more battery information.
CAUTION
When charging the batteries, ensure the room temperature is between 15° C
(59° F) to 30° C (86° F) to minimize charge time and maximize battery life.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...