COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
9-5
Technical Specifications
Recommended Practices
COLLEAGUE Communication Cable
The COLLEAGUE Communication Cable (Baxter part number
AS3IS3002) is used with COLLEAGUE GUARDIAN Configuration
Tool and the COLLEAGUE DL2 Event History Download Application,
allowing for transfer of information between COLLEAGUE pumps and
PCs. The cable is attached to the Communications Port located on the
rear of the pump (Figure 2-3). See “Optional Pump Accessories,” 5-1 for
more information on COLLEAGUE GUARDIAN Configuration Tool
and the COLLEAGUE DL2 Event History Download Application.
The pump should not be connected to the patient when the
COLLEAGUE Communication Cable is attached; it is not suitable for
use in the patient environment.
External Monitoring
The External Monitoring feature is for Baxter diagnostic purposes only.
Recommended Practices
Connections of this pump into the same patient line with other
infusion systems or accessories may alter the system
performance. Consult the infusion system or accessory
manufacturer’s instructions for use before proceeding.
To ensure that pump performance is maintained, annual
inspections should be performed by Baxter-trained, qualified
personnel in accordance with the
COLLEAGUE Pump Global
Service Manual
.
In the U.S., annual inspections should be performed in
accordance with The Joint Commission procedures.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...