4
Secondary Infusions
4-46
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Preparing a Secondary Infusion Set
1.
Prepare solution containers and administration sets.
2.
Lower the primary container using the hanger provided with the
secondary set. Ensure the hanger is fully extended.
3.
Load the CONTINU-FLO
administration set into the desired pump
channel as described in “Loading the Administration Set,” 4-16.
4.
If desired, program the primary infusion for the selected channel.
Programming a Secondary Rate-Volume Infusion
1.
From the Main Display screen:
! WARNING !
! WARNING !
! WARNING !
Failure to open the roller clamp on a secondary set when starting a secondary
infusion will cause a delay of the secondary infusion and an over-infusion of the
primary infusion.
For single channel pumps:
For triple channel pumps:
press the
Secondary
soft key.
press the desired
Channel
Select
key, then press the
Secondary
soft key.