Programming a COLLEAGUE GUARDIAN Dose Mode Infusion
11-12
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
6.
(Optional) If the label is set up to allow non-standard concentration
programming, the drug amount, diluent volume, and concentration
can be changed by using the
×Ø
keys to highlight the appropriate
field and entering new values using the numeric keypad.
7.
Use the
Ø
key to highlight the
Dose
field and enter the dose. The
pump displays the dose and calculated flow rate.
If values are changed so that the resulting drug amount, diluent
volume, or concentration is non-standard, the changed values are
indicated by white triangles beside them (Figure 11-12).
8.
Press the
Confirm Primary
soft key.
If the dose entered is outside the dose limits, a Limits Warning
pop-up is displayed (Figure 11-13). Do one of the following:
Press
Cancel Dose
(
Ø
key) to cancel the dose, then enter a dose
that is within the preset limits.
If the clinical decision is to proceed with the override of the
COLLEAGUE GUARDIAN limits, press
Accept Dose
(
×
key)
to accept the out-of limits dose and continue.
Figure 11-12 Non-Standard
Programming (White Triangles)
Figure 11-13 Limits Warning
Pop-up
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...