COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
4-21
Operating Instructions
Primary Infusions
Programming an Infusion
Before starting an infusion, the administration set must be fully loaded.
If an administration set has not been loaded and
START
is
pressed, a
Tube Not Loaded
alarm will occur.
If an administration set is still loading and
START
is pressed, a
Tube Loading in Progress
alarm will occur.
See Table 8-1, “Troubleshooting Alarm Messages,” on page 8-13 for
more information about these alarms.
Primary Infusions
Programming a Primary Rate-Volume Infusion
1.
From the Main Display screen (Figure 4-24):
The display then changes to the Rate-Volume programming screen
(Figure 4-25). The
Rate
field is highlighted.
Figure 4-24 Main Display Screen
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key or
the
Rate
or
Volume
keys.
press the desired
Channel
Select
key.
Note:
Channel Stopped
Alert
If a pump channel is powered on and no keys are
pressed for two minutes, a
Channel Stopped
alert
occurs. Either continue programming, start the infu-
sion, or place the channel in standby to clear the alert.
! WARNING !
There may be periods of no flow for flow rates less than
or equal to 1mL/hr.
Summary of Contents for Colleague 3 CXE
Page 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Page 229: ...071960491 ...