15
en
7.4
Failure messages
A failure is indicated as follows:
■
Clear sound sequence (as necessary, may interrupt the sound of the HF
discharge)
■
The red error light
6
on the front side of the HF generator lights up
■
Error message in the display
5
on the front side of the HF generator
Note
If the failure occurs during HF discharge, the vessel sealing may be incom-
plete or insufficient.
In order to ensure that sufficient hemostasis has been achieved, follow the
procedure for regrasp errors, see Regrasp error.
Note
If the failure continues after turning the device off and on again twice, con-
tact your national B. Braun/Aesculap agency, see Technical Service.
Failure code
Cause
Remedy
F001
Software error on the HF generator
Switch off HF generator, wait 5 s, switch HF generator on
F002
Dosage error of HF generator
Switch off HF generator, wait 5 s, switch HF generator on
F003
Hardware error
Switch off HF generator, wait 5 s, switch HF generator on
F004
Internal communication error of the HF generator
Switch off HF generator, wait 5 s, switch HF generator on
F005
Overheating of HF generator
Switch off HF generator and leave to cool
Check the air stream around the generator
F006
Error in the operating mode selection button (MODE)
2
(activated
when switching on)
Switch off HF generator, wait 5 s, release operating mode
selection button (MODE)
2
, switch HF generator on
F007
Error with HF energy output
Switch off HF generator, wait 5 s, switch HF generator on
F008
Fault: HF outside the calibration
Switch off HF generator, wait 5 s, switch HF generator on
F009
Fault: Internal power supply
Switch off HF generator, wait 5 s, switch HF generator on