16
en
7.5
Problem solving
7.6
User maintenance
►
In case of error, follow the instructions in the display and note the error
code.
►
If the error continues after following the instructions in the display,
contact your national B. Braun/Aesculap agency, see Technical Service.
7.7
Fuse replacement
Specified fuses: 2 units Aesculap art. no. TA021404:
G fuse link, time lag (T) 8 A, breaking capacity H (1 500 A)
►
Use a small screwdriver to release the clip on the fuse holder
11
and
tilt up cover.
►
Use a small screwdriver to release the fuse holder
11
and pull it out.
►
Replace both fuses.
►
Reinsert the fuse holder
11
.
►
Close the cover.
Fig. 14
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
8.
Technical Service
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
Other service addresses can be obtained from the address indicated above.
9.
Accessories/Spare parts
Finding
Remedy
HF generator does not switch on
Ensure that the power cord has been correctly plugged into the power supply
13
of the HF gen-
erator.
Ensure that the power cord is connected to a functioning power source.
Ensure that the power OFF/ON switch
12
of the HF generator is in the power ON position.
Check the fuse in the fuse box and replace if necessary, see Fuse replacement.
If the error continues, replace power cord.
The device is switched on, but no displays are illumi-
nated and the self-test is not being carried out
Switch off HF generator, wait 5 s, switch HF generator on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Service.
Self-test fails
Switch off HF generator, wait 5 s, switch HF generator on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Service.
HF generator is active (power ON signal lamp
9
lights
up), and accessories are connected, but there is no HF
discharge
Verify that the required foot switch (if used ) is properly connected and plugged in.
Ensure that the Caiman Seal & Cut Instrument is correctly connected and plugged in (green
ring
1
is lit up).
In order to start the HF discharge, activate the button on the instrument or the foot control.
If the error continues, replace Caiman instrument.
DANGER
Risk of death by electric shock!
►
Unplug the device before changing the fuses!
DANGER
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
►
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
►
Do not modify the product.
Art. no.
Designation
TA021404
Fuse
–
Caiman 5 instrument family
–
Caiman 12 instrument family
GN201
Foot switch