17
en
Note
Other accessories, applied parts and replacement parts are described in
Aesculap brochure C-304-81 and C-902-02
10. Technical data
Classification acc. to Directive 93/42/EEC
10.1
Ambient conditions
Note
The atmospheric pressure of 700 hPa corresponds to a maximum operating
altitude of 3 000 m.
Note
After transporting or storing at temperature outside the operating temper-
ature range, leave the HF generator for one hour before use so that it can
reach room temperature.
11. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
►
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
Art. no.
Designation
Class
GN200
High frequency surgical device
IIb
Mains voltage ranges
(Current intake)
100V~–240 V~ (6.7 A–2.8 A)
Frequency
50–60 Hz
Protection class (acc. to
IEC / DIN EN 60601-1)
I
Foot switch circuit
ignition-safe acc. to IEC/DIN EN 60601,
approved for use in “medical environment”
Output power
2 x 150 W on 50 Ohm
Device protection
Time-Current characteris-
tic
Breaking capacity
Construction
T8 AH/250 V
T (Time lag)
H (1 500 A)
5 mm x 20 mm
HF operating frequency
460 kHz ± 1 %, Quasi-Sinus
Operating mode
Int 10 s/30 s
Weight
8.1 kg
Dimensions (L x W x H)
400 mm x 355 mm x 100 mm
Application part type acc.
to IEC/DIN 60601-1
CF
Recovery time of the defi-
brillation-protected out-
put
0 seconds
CISPR11
Class A
EMC
IEC/DIN EN 60601-1-2
Conforming to standard
IEC/DIN EN 60601-1
IEC/DIN EN 60601-2-2
Operation
Storage and transport
Temperature
Relative
humidity
non-condensing
non-condensing
Atmospheric
pressure
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
-10
-10
50
10
90