UR14 Driver/Receiver Board I-27
Note: Use the UR14 Current Estimator calculation tool to estimate the power
converter power consumption from the ±12V and ±24V power rails. This tool
is available upon request from Astronics Test Systems at
Environmental
Table I-11: Environmental
Temperature
Operating: 0° C – 45° C
Storage: -40° C – 70° C
Humidity
5% to 95%
Altitude
10,000 ft
Cooling Required
(10°C Rise; 2 UR14s)
Max: 14.4 lps @ 2.8 mmH
2
0
Typ.: 10.4 lps @ 1.6 mmH
2
0
MTBF (ground benign)
UR14: 179,889 hours
T940: 180,855 hours
Power Converter: 540,040 hours
T940-UR14: 77,279 hours
Dimensions
20 x 114 x 305 mm
EMC (Council Directive
89/336/EEC)
Emission: EN61326-1: 2006, Class A
Immunity: EN61326-1: 2006, Table 1
Designed to Meet – Testing in Progress
Safety (Low Voltage Directive
73/23/EEC)
BS EN61010-1: 2007