Publication No. 980938 Rev. K
Model T940 User Manual
Astronics Test Systems
Programmable Channel Calibration 6-21
2. Verify that the ISource/ISink
calibrate function is now in focus.
Select Start and End Channels and Measurement Delay
Equipment: Basic Setup
Procedure:
1. Select DRA or DRB (if installed) using the Driver/Receiver switch.
2. Use the Start and End Channel fields to select the I/O and Auxiliary
channels to be calibrated.
3. The default measurement delay is 200 ms. Increase this value to give
the calibration points more time to settle.
DRM Calibration Warmup
Equipment: Basic Setup
Procedure:
1. Allow the T940 DRM to warm to its nominal application temperature.
2. Hit the Continue button when the required temperature is reached. If
the temperature reaches 80°C, the process continues automatically.