Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-122
Astronics Test Systems
outputs. The sync parameters consist of an offset and a length. Once the
programmed sync event occurs, the sync pulse will begin after the "offset" and
last for "length". Both "offset" and "length" are specified in pattern clocks. The
sync pulse will not extend past the end of the sequence. In the "Step" event, the
sync pulse will not extend beyond the specified step.
Figure 5-64: Set Sync Panel
Sync Number
This control selects which sync pulse signal to program, either Sync 1 or Sync 2.
Event (and Step)
This pull-down control programs the sync event. The sync pulse event can be
set to either the start of a sequence or a specific step.
The selections for this pull-down control are:
Table 5-91: Finish Mode Settings
Setting
Description
Start
The sync pulse begins from the start of the
sequence.
Step
The sync pulse begins from the specified step.
The relevant VXI
plug&play
API function is:
•
tat964_setSyncEvent
Offset
This control sets the offset from the sync event before the sync pulse starts. The
offset can be set from 0 to 1048575 patterns.
The relevant VXI
plug&play
API function is:
•
tat964_setSyncParameters