Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-156
Astronics Test Systems
Channel
This control sets the channel number to measure.
The relevant VXIplug&play API functions are:
•
tat964_pmuMeasureVoltage
Measure Voltage
This control initiates a voltage measurement.
The relevant VXI
plug&play
API function is:
•
tat964_pmuMeasureVoltage
Instrument Functions
The Instrument menu bar selections allow the user to perform the following:
•
Self-test functions
•
Calibration
•
Firmware Updates
•
Temperature Monitoring
•
Voltage Monitoring
The instrument functions are dependent on the Driver/Receiver boards installed.
For example; the DR1 Driver/Receiver board does not require voltage calibration
and does not contain voltage and temperature monitoring hardware.
Self Test
The self-test function is accessed from the
Instrument >Self Test
menu bar
selection.
Figure 5-86: Self Test Result Message
The self-test function resets the instrument and performs a short RAM test on all
the internal memories. The short RAM test tests each RAM at the major address
bits locations, i.e., 0, 1, 2, 4, 8, 16 . . . etc.
The self-test result codes are: