Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-38
Astronics Test Systems
Figure 5-25: Data Sequencer Configure Settings Panel
Error Record Basis
This pull-down control programs the sequencer error record basis.
This control allows the user to select how the response data will be evaluated for
errors when the record mode is set to
Record Errors
.
The selections for this pull-down control are:
Table 5-34: Error Record Basis Settings
Setting
Description
Dual
Use both good 1 and good 0 comparator
levels
Good 1
Use only the good 1 comparator (Single
threshold)
These two choices are provided for use with the DR3e Driver/Receiver only.
The relevant VXI
plug&play
API function is:
•
tat964_setRecordParameters
Raw Record Basis
This pull-down control programs the sequencer raw record basis.
This control allows the user to select which comparator will be used to determine
the data level when the record mode is set to
Record Response
.
The selections for this pull-down control are:
Table 5-35: Raw Record Basis Settings
Setting
Description
Good 0
Use good 0 comparator levels (only available
on dual threshold Driver/Receiver boards like
the DR3e).
Note: The Good 0 is complemented when
recorded
Good 1
Use good 1 comparator levels
These two choices are provided for use with the DR3e Driver/Receiver.
The relevant VXI
plug&play
API function is:
•
tat964_setRecordParameters
Record Offset
The record offset allows the user to shift the record signals (pattern code expect
and mask, record offset, window strobes) to accommodate system and UUT