Publication No. 980938 Rev. K
Model T940 User Manual
Astronics Test Systems
Soft Front Panel Operation 5-159
Figure 5-88: Power Converter Test Results Panel
The power converter test saves the current power converter setting and performs
a test on all the power converter modes. The power converter test verifies that
the positive and negative rails are within 3% of nominal. The min/max thresholds
for each mode are listed in the following table:
Table 5-107: Power Converter Test Thresholds
Power
Converter
Mode
V+ Min
V+ Max
V- Min
V- Max
-12 to +12
15.520
16.480
-16.068
-15.132
-15 to +5
9.312
9.888
-19.776
-18.624
-10 to +10
15.520
16.480
-14.523
-13.677
-2 to +7
11.640
12.360
-9.888
-9.312
-5 to +15
18.624
19.776
-9.888
-9.312
0 to +24
27.936
29.664
-4.635
-4.365
-2 to +22
25.608
27.192
-6.180
-5.820
The relevant VXIplug&play API functions are:
•
tat964_setPowerConverter
•
tat964_setPowerConverterState
•
tat964_queryAdcAverage
Calibration Panel
The calibration function is accessed from the
Instrument >Calibrate
menu bar
selection. Reference Chapter 6 Programmable Channel Calibration for field
calibration procedure.
Calibration data is stored on the Driver/Receiver board in non-volatile memory.