Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-104
Astronics Test Systems
Setting
Description
Step FAIL
Jump if the PASS/FAIL flag is equal to FAIL
Step PASS
Jump if the PASS/FAIL flag is equal to PASS
Sequence FAIL
Jump if Burst Error Count is not equal to zero.
Sequence PASS
Jump if Burst Error Count is equal to zero.
Jump Trigger 1
True
Jump if “Jump Trigger 1” true.
Jump Trigger 1
not True
Jump if “Jump Trigger 1” not true.
Jump Trigger 2
True
Jump if “Jump Trigger 2” true.
Jump Trigger 2
not True
Jump if “Jump Trigger 2” not true.
Jump Trigger 3
True
Jump if “Jump Trigger 3” true.
Jump Trigger 3
not True
Jump if “Jump Trigger 3” not true.
Jump Trigger 4
True
Jump if “Jump Trigger 4” true.
Jump Trigger 4
not True
Jump if “Jump Trigger 4” not true.
The true/false state of the jump triggers is based on the jump trigger test
condition. If the jump trigger test condition is set to “Low Level”, then “True”
would indicate the jump trigger signal is low and “not True” would indicate the
jump trigger signal is high.
Note: Any CONDEN enabled FAIL during the Sequence Step will prevent a
PASS.
See the
Jumping, Halting, Counting and Logging Errors
section of
Chapter
8
for a detailed explanation of Jumping based on Errors. Also see the
Jumping on
a Step or Burst Error
section of
Chapter 8
.
The relevant VXI
plug&play
API function is:
•
tat964_setSequenceJump
Loop Count
This numeric control programs the Loop Count number.
Jumps can be qualified by a loop counter. The loop count can be set from 0 (no
qualification) to 65536. A count qualified jump only allows the jump to occur a
maximum of “count” times. This allows single or multiple steps to be looped.
The relevant VXI
plug&play
API function is:
•
tat964_setSequenceJump