Model T940 User Manual
Publication No. 980938 Rev. K
Programmable Channel Calibration 6-6
Astronics Test Systems
2. Verify that the ADC Reference calibrate function is now in
focus.
Select Measurement Delay
Equipment: Basic Setup Procedure:
1. Select DRA or DRB (if installed) using the Driver/Receiver switch.
2. The default measurement delay is 200 ms. Increase this value to give
the calibration points more time to settle.