86
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 7
Temperature non-uniformity test
8. Slide the heated cover back, and repeat
and
of this procedure.
Complete these steps for all 12 wells to be tested:
A1, A6, A10, B3, C5, C10, D1, D7, E8, F1, F5, and F10
9. The GeneAmp
®
9700 evaluates the uniformity of the sample block temperature
for the setpoint values that you entered and displays the results. A summary
screen opens at the end of the test.
If you entered values on the Temperature Non-Uniformity Test Data Sheet,
compare those values with the actual test results.
10. Press
F1
(Accept). To interpret the results, see
.
11. When you complete all measurements, be sure to:
• Press
Cancel
.
• Remove the 0.5-mL probe assembly from the sample block.
• Power off the digital thermometer, then clean off the oil.
• Remove the 0.5-mL temperature verification frame from the sample block.
IMPORTANT!
Make sure the sample block is at room temperature (~25 °C)
before removing the frame.