Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 8
Data Sheet: Calibration Verification Test
Data Sheet: Calibration Verification Test
When running the calibration verification test, record the setpoint values for well A6
on this data sheet. At the end of the test, compare the values that are displayed on the
thermal cycler with the values that are recorded here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial no.
Setpoint value: Well A6
85 °C
45 °C