38
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Calibration verification test
7. For GeneAmp
®
9700 with Manual Dual 96-Well Sample Block Module:
Slide the heated cover forward then pull the lever down.
For GeneAmp
®
9700 with Auto-Lid Dual 96-Well Sample Block Module:
Press the red button to slide and close the heated cover.
IMPORTANT!
Seat the probe properly and close the heated cover carefully. If
the probe wire is crushed when you close the heated cover, the probe can be
damaged.
8. Power on the digital thermometer.
Refer to the instructions in your Temperature Verification Kit for details on
operating the digital thermometer, Model 4500.
Configuring the
GeneAmp
®
9700
1. Power on the thermal cycler. The Main menu opens.
For best results, allow the thermal cycler 15 to 30 minutes to warm up to
ambient temperature before performing the test.
2. Press
F4
(Util).
The Utilities screen opens.
3. Press
F1
(Diag). The Diagnostics screen opens.
4. Press
F3
(TmpVer). The Temperature Verification screen opens.
5. Press
F1
(Temp).
The GeneAmp
®
9700 is automatically configured for the calibration
verification test, and the Calibration Verification screen opens.
Temperature Verification
Temp - Calibration Verification
TNU - Temperature Non-Uniformity
F1
F3
F4
F5
F2
Temp
Exit
TNU
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Place probe in well R3
Press Run
F1
F3
F4
F5
F2
Run
Cancel