Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 7
Data Sheet: Temperature Non-Uniformity Test
Data Sheet: Temperature Non-Uniformity Test
When running the temperature non-uniformity test, record the setpoint values for the
wells listed on this data sheet. At the end of the test, compare the values that are
displayed on the thermal cycler with the values that are recorded here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial no.
Setpoint value
94 °C
37 °C
A1
A6
A10
B3
C5
C10
D1
D7
E8
F1
F5
F10