Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 2
Temperature calibration verification test
3. Press
F1
(Diag). The Diagnostics screen opens.
4. Press
F3
(TmpVer). The Temperature Verification screen opens.
5. Press
F1
(Temp). The thermal cycler is automatically configured for the
calibration verification test and the Calibration Verification screen opens.
Running the test
In this test, you take temperature readings of the sample well at two different setpoint
temperatures using the 0.2-mL probe assembly.
Note:
Press
F5
(Cancel) if you want to end the test.
To run the calibration verification test:
1. Press
F1
(Run) to start the calibration verification test. The Calibration
Verification screen opens with the setpoint value displayed.
Note:
The cover must be at 85 °C
±
10 °C. It may take several minutes for the
system to ramp up.
Note:
The Calibration Verification screen counts down the time until the
setpoint is reached.
Temperature Verification
Temp - Calibration Verification
TNU - Temperature Non-Uniformity
F1
F3
F4
F5
F2
Temp
Exit
TNU
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Place probe in well A6
Press Run
F1
F3
F4
F5
F2
Run
Cancel
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Setpoint is 85°C
Cover must be within 10°C of 85°C
F1
F3
F4
F5
F2
Cancel
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Stabilizing at setpoint
x:xx
F1
F3
F4
F5
F2
Cancel