![Applied Biosystems 2720 User Manual Download Page 106](http://html.mh-extra.com/html/applied-biosystems/2720/2720_user-manual_2955102106.webp)
98
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 8
Temperature non-uniformity test
Configuring the
GeneAmp
®
9800
“Setting up the 0.1-mL probe assembly” on
.
2. Power on the GeneAmp
®
9800.
For best results, allow the thermal cycler 15 to 30 minutes to warm up to
ambient temperature before performing the test.
3. In the main menu, press
F4
(Util)
to open the Utilities screen.
4. Press
F1
(Diag) to open the Diagnostics screen.
5. Press
F3
(TmpVer) to open the Temperature Verification screen.
6. Press
F2
(TNU) to open the TNU Performance screen and configure the 9800
for the temperature non-uniformity test.
Note:
If the cover temperature is not within
±
1 °C of the heated cover setpoint,
the following screen opens:
Temperature Verification
Temp - Calibration Verification
TNU - Temperature Non-Uniformity
F1
F3
F4
F5
F2
Temp
Exit
TNU
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Place Probe in well A1
Press Run
F1
F3
F4
F5
F2
Run
Cancel
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Cover setpoint is 55C
Cover must be within 1.0C of setpoint
F1
F3
F4
F5
F2
Cancel