40
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Calibration verification test
4. Press
Enter
. The GeneAmp
®
9700 automatically begins the second reading
(45 °C
setpoint), and the Calibration Verification screen opens with the setpoint
value displayed.
Note:
The cover must be at 105 °C
±
1 °C. It may take several minutes for the
system to ramp up.
for the second temperature reading.
The GeneAmp
®
9700 evaluates the calibration of the sample block temperature
for the setpoint values that you entered and displays the results. A summary
screen opens at the end of the test.
6. If you entered values on the Calibration Verification Test Data Sheet (
compare those values with the actual test results.
7. Press
F1
(Accept).
Calibration Verification
Block temp = xx.x°C Cover temp = xxx°C
Setpoint is 45°C
Cover must be within 1°C of 105°C
F1
F3
F4
F5
F2
Cancel
Calibration Verification
Actual temperature at 85
°
C
Actual temperature at 45
°
C
F1
F3
F4
F5
F2
Accept
Cancel
xx.x
xx.x