Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 10
Data Sheet: Calibration Verification Test
Data Sheet: Calibration Verification Test
When you run the calibration verification test, record the setpoint values for well
E1
on this data sheet. At the end of the test, compare the values that are displayed on the
thermal cycler with the values that are recorded here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial no.
Avg Probe Temp 1 at
85 °C
Avg Probe Temp 2 at
45 °C
Left
Zone
Right
Zone
Left
Zone
Right
Zone