122
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 10
Calibration verification test
Ending the test
1. If the heated cover is closed, lift the lever and place the heated cover in the fully
opened position.
PHYSICAL INJURY HAZARD.
Hot Surface. Use care
when working around the heated cover and sample block to avoid being burned
by hot components.
2. Wait for the sample block to reach room temperature (~25 °C), then remove the
the 0.5-mL probe assembly and thermal isolation frame from the sample block.
3. If you are performing a repeat measurement, clean the probe assembly and all
wells of mineral oil, then perform the repeat measurement. Otherwise, power off
the Veriti
®
Thermal Cycler and Model 4500 digital thermometer.
4. Clean the oil off the sample block.
Note:
Follow the decontamination procedures in user guide for cleaning oiled
wells.