Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 9
Data Sheet: Temperature Non-Uniformity Test
Data Sheet: Temperature Non-Uniformity Test
When you run the temperature non-uniformity test, record the setpoint values for the
wells listed on this data sheet. At the end of the test, compare the values that are
displayed on the thermal cycler with the values that are recorded here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial no.
Setpoint value
95 °C
40 °C
A1
A4
A8
A12
C1
C4
C8
C12
E1
E4
E8
E12
H1
H4
H8
H12