104
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 8
Data Sheet: Temperature Non-Uniformity Test
Data Sheet: Temperature Non-Uniformity Test
When running the temperature non-uniformity test, record the setpoint values for
wells A1, A12, C4, C9, F4, F9, H1, and H12 on this data sheet. At the end of the test,
compare the values that are displayed on the thermal cycler with the values that are
recorded here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial no.
Setpoint value
95 °C
55 °C
A1
A12
C4
C9
F4
F9
H1
H12