Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Overview
Overview
Materials
When you perform the temperature verification test on a GeneAmp
®
9700 with a
dual 96-well sample block module, you use the materials in the kit (PN 4317939).
Assembly
show the right frame of a properly assembled temperature verification
system. Note that the probe wire is threaded through the channel to prevent damage
to the wire when you close the thermal cycler heated cover.
Figure 7
The temperature verification system for GeneAmp
®
9700 with a dual
96-well sample block module (right frame shown)
Figure 8
Left frame
Figure 9
Right frame