q
Manual
— Sets the input attenuator manually from 0 to 51 dB. Choose a
value that does not overload the instrument.
q
Dynamic
— Sets the value of the input attenuator so that it is dynamically
coupled to the input signal and turns the preamp on or off based on the
overall signal conditions across the entire frequency range of the instru-
ment. The instrument will set the lowest input attenuation consistent with
the need to not saturate the front end.
NOTE: A large signal that is not displayed on-screen can sometimes cause the
instrument to turn off the preamplifier or add input attenuation, causing the
noise floor to jump up or down suddenly as the large signal turns on or off.
q
Preamp Control Manual
— Activates the preamp menu for manual control
of the preamplifier.
q
Preamp On/Off
— Sets the preamplifier on or off.
q
Preamp Auto
— Automatically adjusts the preamplifier according to
the reference level.
q
Back
— Returns to the previous menu level.
q
Units
— Choose from the menu of amplitude related units. When Linear is
selected, the available selections are Volts or Watts. When Log is selected,
the selection of dBm sets absolute decibels relative to 1 mW as the amplitude
unit. Selection of dBV, dBmV or dB
m
V sets absolute decibels relative to 1
volt, 1 millivolt, or 1 microvolt respectively as the amplitude unit.
q
Ref Level Offset
— Sets the reference level offset, which is a value, in dB, of
gain or loss external to the instrument. Negative values represent external at-
tenuation. This feature allows measurement of high gain devices in combina-
tion with an attenuator, and is used to offset the reference level to view the
correct output level. For example, to measure a high gain amplifier with an
output of 70 dBm, an external 50 dB attenuator must be inserted between the
instrument and the device. To compensate, set the reference level offset to
–50 dB.
MEAS/DISP
Pressing
MEAS/DISP
activates a menu of bandwidth and sweep-related func-
tions. Use the corresponding soft key to select the desired function.
q
Spectrum
— Spectrum selects a traditional spectrum analyzer display.
q
Spectrogram
— Selecting Spectrogram provides a three dimensional repre-
sentation of frequency, time and power to aid in the identification of intermit-
tent signals. The bottom portion of the display shows a traditional spectrum
analyzer plot while the upper portion shows the power level in color with the
most recent trace placed at the bottom of the display. A color scale on the left
side of the display translates color into power level. The highest power level
is displayed in red with the lowest power levels in dark blue and black.
q
Sweep Interval
— Sweep Interval allows for the amount of time between
sweeps to be set from a minimum of zero seconds up to 660 seconds (11
minutes). Enter the number using the Up/Down arrow key or use the key-
pad and select the appropriate soft key for minutes, seconds, milliseconds
or microseconds. The current setting will be displayed in the message area.
A sweep interval of 0 seconds will display as
Auto
and the instrument will
sweep as fast as possible. Setting this value also determines the time span.
2-52
Chapter 2 Functions and Operations
Summary of Contents for Cell Master MT8212B
Page 3: ......
Page 5: ......
Page 222: ...12 4 Chapter 12 GPS Feature Figure 12 3 Saved GPS Location Information...
Page 231: ...Chapter 13 CDMA Measurement 13 9 Figure 13 5 CDMA Display with Markers...
Page 255: ...Chapter 18 EVDO Mode 18 5...