Dynamic Attenuation Control
The dynamic attenuation control feature tracks the peak input signal level and automatically
adjusts the attenuator to protect the MT8212B in situations of high RF signal levels. This
feature also enhances the instrument sensitivity in situations of low-level RF signals by de-
activating and activating the internal preamplifier resulting in improved measurement accu-
racy.
When dynamic attenuation is on, the MT8212B is in control of the attenuation and preamp.
Manual settings will be overwritten based on the input signal. To manually control these
parameters, set the attenuation to
Manual
.
The following is an example identifying interference in a PCS 900 MHz band with the aid
of the preamplifier and dynamic attenuation control.
Required Equipment
q
MT8212B Handheld Cell Master
q
A PCS band Pass Filter, such as K & L Electronics 5DR30-921 or equivalent
q
A PCS band antenna
Procedure
Step 1.
If necessary, press the
MODE
key and use the Up/Down arrow key to select the
Spectrum Analyzer
mode. Press the
ENTER
key to select.
Step 2.
Reset the MT8212B to the factory default settings for the current mode by press-
ing the
RECALL SETUP
key on the data keypad. When the key is pressed, a
Recall Setup selection box appears on the display. Select setup location 0 <De-
fault> to recall the factory preset state for the currently selected mode.
Step 3.
Connect the PCS antenna to the input of the band pass filter, and the output of
the band pass filter to the
RF In
connector of the MT8212B.
Step 4.
Press the
FREQ/DIST
key.
Step 5.
Press the
Start
soft key and enter 1.93 GHz.
Step 6.
Press the
Stop
soft key and enter 1.935 GHz.
Step 7.
Press the
AMPLITUDE
key and select the
Scale
key. Set the scale to 10 dB/di-
vision.
Step 8.
Press the
Atten/Preamp
soft key and set the attenuation to
Dynamic
.
As the MT8212B sweeps, it continually monitors the overall input signal level and, as re-
quired, adjusts the input attenuation value and the preamplifier on/off state to achieve the
best possible sensitivity.
5-6
Chapter 5 Spectrum Analyzer Measurements