Chapter 17
Signal Generator Mode
Introduction
This chapter presents information and procedures to make measurements using the optional
CW signal generator mode (Option 28). The CW Signal Generator provides a CW signal
from the
RF Out
port of the instrument. The CW signal is primarily used for sensitivity test-
ing of receivers. To test receiver sensitivity, connect the signal directly to the receiver being
measured, and then reduce the output amplitude until the receiver drops the signal. The ex-
ternal splitter feeds the signal into the RF input of the instrument. The display shows the
output power and frequency.
Required Equipment
q
Cell Master MT8212B with Option 28
q
Attenuator
q
Splitter
Procedure
Step 1.
On the Cell Master, press the
MODE
key, use the Up/Down arrow key to select
the
CW Signal Generator
mode and press
ENTER
.
Step 2.
Press the
Freq
soft key, use the Up/Down arrow key to change the desired fre-
quency, or use the keypad to enter the desired frequency and select the
MHz
soft
key.
Step 3.
Connect the attenuator to the
RF Out
port and the splitter to the
RF IN
port.
17-1
ATTENUATOR
SPLITTER
TO RECEIVER
Figure 17-1.
CW Signal Generator Mode Setup