Step 7.
Press the
ENTER
key.
Step 8.
Disconnect the 18" (46cm) Test Port Extension Cable from the
RF In
connector
and connect the phase stable cable and attenuator between the
RF In
connector
and the 18" (46cm) Test Port Extension Cable.
Step 9.
Press the
ENTER
key.
Step 10.
A message will appear confirming that the Offset Calibration is done and show-
ing the offset value to be applied. Press the
ENTER
key to accept the offset
value, or press
ESCAPE
to cancel. The message
Offset Cal On (# dB)
will show
at the bottom of the display.
6-3
Test Port Extension Cable
15NNF50 – 1.5C
30 dB Attenuator
42N50A-30
18" Cable
3-806-151
10 dB Attenuator
3-1010-119
RF Out
RF In
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT
MARKER
SAVE
DISPLAY
RECALL
DISPLAY
MODE
FREQ/DIST
AMPLITUDE
MEAS/DISP
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
/
1
2
4
5
6
7
8
9
0
3
+-
.
MT8212B
Cell
Master
Figure 6-2.
Measuring Insertion Loss of the Test Port Extension Cable and 30 dB Attenuator