3-55
Burst Size pane
Burst Sizes can be set in two modes: Stepped or Constant. Use the Burst
Mode field to select the mode.
3
Ether
n
et Applicat
ion
Figure 3.8.3-2 RFC2544 Burst Setup Screen, Burst Size Pane in
Stepped Mode
In the Stepped mode, the test burst size range is specified by Start burst
size and End burst size. The burst size is varied in intervals with the
value specified in the Step burst size field.
Figure 3.8.3-3 RFC2544 Burst Setup Screen, Burst Size Pane in
Constant Mode
In Constant mode, only one burst size is used for each frame size. The
burst size is specified in the Constant burst size field.
Advanced pane
The advanced setup is similar to that of the RFC2544 Throughput test
setup. See section 3.8.1 for further details.