4-28 | Thermox CEM O2/TM
Overview of Alarm Functions
Figure 4-7 provides an overview of how alarms can be set:
Alarms 1 and 2 are reserved for the watchdog alarm (Alarm 1) and the
service alarm (Alarm 2). These alarms are not software selectable.
The Select Function menu option applies only to Alarm 3.
The Relay Configure menu option applies to all alarms except for the
watchdog and service alarms, which always de-energize on alarm (fail
safe).
A
lARM NUM
b
ER
FUNCTION
*Select Function
Alarm Configure
Relay Configure
Alarm 3
% O
2
Cal Start
Verify Start
High O
2
Alarm
Low O
2
Alarm
(Only selectable if
Select Function is
oxygen)
(Disabled during
calibration cycle and
during system alarm
condition)
Energize on alarm
De-energize on alarm
Alarm 4
O
2
Hydrocarbon Upset
(Selectable through
Setup/Sensor Con-
fig. menu option)
High O
2
Alarm
Low O
2
Alarm
(Disabled during
calibration cycle and
during system alarm
condition)
* Note: Alarms 1 and 2 are reserved for the watchdog timer and service alarms and are
not software-selectable.
Figure 4-7.
Alarm Key overview.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...