Calibration | 5-25
Configure Verify
The Configure Verify menu option allows you to define the verify gas
value and the acceptable range of readings for a verify.
To set your verify gas value and acceptable verify range, do the following:
1. Select Configure Verify from the Calibrate Key menu.
2. Select whether you wish to enter the verify gas using a percent or
PPM scale.
You will be prompted to enter the verify gas value:
Verify Gas = xx O2
New Value?
3. Enter the verify gas value, then press the
Enter
key. If you wish to
keep the current verify gas value unchanged, press the
Enter
key
without entering a value.
4. Enter the acceptable verify range. For example, if you enter a verify
gas range of 1%, and the verify gas value is 10%, then the acceptable
range is between 11% and 9%.
If during a verify operation (Initiate Verify menu option) the system
finds that the verify gas is not within the acceptable range, a Verify
Failure message will appear on the bottom line of the display, and
the Service Alarm will trip. The error message will continue to be
displayed until a successful verify is completed (gas value within the
range) or a successful calibration is completed.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...