Calibration | 5-17
Inject Cal Gas
This menu option requires that an auto cal card and RCU are in-
stalled.
This menu option allows you to manually adjust the pressure and flow
of your calibration gas (pressure of calibration gas cylinders and flow into
sensor) by cycling through the solenoid valves on the RCU. No calibration
or verification data is collected while in this menu option. See the “Me-
chanical Installation” section in the Installation chapter for recommended
calibration gas pressure and flow requirements.
It is important that all calibration gases, as well as the sample gas, flow
through the analyzer at the same flow rate.
1. Connect the span and zero gas cylinders to the RCU.
2. Select the
Inject Cal Gas
menu option from the Calibrate key menu.
The following message will then appear on your display, indicating that
the RCU has introduced the span gas to the sensor:
Span/Verify gas is flowing
You will be prompted to press the
Enter
key once you have manually set
the pressure and flow for the span gas:
ENTER for next Gas
CANCEL to quit
3. When you have adjusted the pressure and flow for the span gas to the
desired levels, press the
Enter
key.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...