5-24 | Thermox CEM O2/TM
Set Cal Timers - Recovery Duration
This menu option applies to all automatic, remote, and manual cali-
brations and verifications.
This menu option allows you to define a recovery time so the control unit
can return to reading process gases after reading calibration gases without
affecting the analog outputs or the triggering of alarms.
If, for example, your process gas is 6%, and your last calibration gas is
2%, this option gives the system time to recover back to reading the 6%
process gas. This menu option also can be used to prevent triggering of
alarms if, for example, your low alarm is set to activate if readings fall
below 3%, and your calibration gas is 2%.
1. Select
Recovery Duration
from the Set Cal Timers menu.
You will be prompted to enter a calibration recovery duration:
Cal Recovery = xx:yy
Enter new time: Mn:Sc
2. Enter the recovery time, then press the
Enter
key. If you wish to use
the currently displayed recovery time, press the
Enter
key without
entering any text.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...