Calibration | 5-9
You are then prompted to apply the O
2
zero gas to the calibration inlet
port of the sensor:
Inject Zero Gas
ENTER to Continue
CANCEL to Abort
6. Press the Enter key once you have applied this cal gas.
The Zero gas display then appears:
O2 xx.x Set xx.x
Press ENTER to Zero
7. Press the Enter key once the zero gas reading on the left of the display
has stabilized. The reading on the far right of the first display line
shows the setpoint value you entered for the O
2
zero gas using the
Cal Gas Values menu option. This O
2
zero gas setpoint value should
match the cal gas cylinder for the O
2
zero gas.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Calibrating
The system then returns to monitoring process readings.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...