Calibration | 5-13
The verify display then appears:
O2 xx.x Set xx.x
Press ENTER to Verify
5. Press the Enter key once the O
2
reading on the far left of the first
display line stabilizes. The reading on the far right of the first display
line shows the set point value you entered using the Configure Verify
menu option. This set point value should match the value on your
verify gas cylinder.
If your O
2
reading on the far left of the display has stabilized, yet is not
near the set point value on the right of the display, check the cal gas value
you entered and ensure that this value matches the cal gas cylinder for
the gas you applied to the cal gas inlet port of the sensor.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Verifying Cal
The system then returns to monitoring process readings.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...