Controller / User Interface | 4-23
Hold during verify
or
Track during verify
After you select one of these options, your choice will be confirmed on the
display:
Out # 1 will TRACK
during Verify
or
Out # 1 will HOLD
during Verify
Set Current Mode Option
The
Set Current Mode
menu option allows you to define either a 0-to-20
or 4-to-20 mA analog output range for the selected analog output port.
This output is dependent on both the range of readings you define (see
the “Set Current Range” section), and on the function you track on the
analog output.
Setting the current mode
1. Select the
Set Current Mode
menu option from the Analog Range Key
after selecting an analog output port.
2. Select
4-to-20 mA Mode
or
0-to-20 mA Mode
.
The following will then appear on the display:
Out # y is set for
4 - 20 ma
or
Out # y is set for
0 - 20 ma
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...