Installation
|
3-21
There is no power switch on the sensor, and it must be protected by in-
stalling it on a circuit protected line, maximum 15 amperes, with a switch
or circuit breaker in close proximity to the sensor and within easy reach of
an operator. Mark the switch or circuit breaker as the sensor disconnecting
device.
Standard Sensor Connections
Make these connections from the control unit to the dry sensor board (see
Figure 3-19):
Figure 3-19.
Standard sensor
connections.
CONTROl UNIT
lAbEl
SENSOR TERMINAl
NUMbER
Cell +
Cell -
1
(one pair)
2
Furnace
15V COM
3
(one pair)
4
T/C +
T/C -
5
(one pair)
6
15V SPLY
15V SPLY
12
(one pair)
12
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...